Here is a list of some of the Scale Calibration Procedures we have collected over the years.  If you cant do it yourself and are in the northeast, give us a call.

A&D

A & D AD 4321

Calibration Procedure for A & D AD 4321 Balance

Turn off Turn throw switch in back press zero CAL1 press stby oper and you should get ................. in display enter the calibration calibration weight using the keys place calibration weight on scale and press stby oper you should get.................. slide switch back take off calibration weight You be done

weight scales

A&D

A&D EK 6000H

Calibration Procedure for A&D EK 6000H Balance

HOLD SAMPLE AND PRINT CAL OUT CAL 0 PRESS SAMPLE SHOWS Cal press print cal 0 prees print add calibration calibration weights press print end remove calibration weight done

weight scales

A&D

A&D EP 60 KB

Calibration Procedure for A&D EP 60 KB Balance

A&D EP 60 KB Throw switch on side cal 0 press print Enter Cal Wt press tare (possibly twice) calF put on calibration weight press tare Cal End

weight scales

A&D

AND FV-30K

Calibration Procedure for AND FV-30K Balance

OPEN BACK OF SCALE PUSH SWITCH TO "ON" PRESS "UNITS" DISPLAY WILL READ "CAL O" PRESS "ZERO" PRESS UNITS DISPLAY READS "CAL 3" WAIT FOR CIRCLE IN UPPER LEFT CORNOR PRESS "ZERO" DISPLAY WILL READ "END" SLIDE SWITCH BACK

weight scales

A&D

AND HV or FW-60K

Calibration Procedure for AND HV or FW-60K Balance

(FV-30K 60LB/40LB)(FW-10K) CAL 1: MAX CAP. IN KG = 150/60/30KG. CAL 2: 2/3 CAP. IN KG = 100/40/20KG. CAL 3: MAX CAP. IN LB = 300/120/60LBS. CAL 4: 2/3 CAP. IN LB = 200/80/40LBS. OPEN UP BATTERY COMPARTMENT REMOVE SCREW FLIP CAL SWITCH TO ON IT WILL READ 9.798 PRESS "MODE/UNITS" "CAL O" STABLE LIGHT PRESS "REZERO" "CAL1" PRESS "MODE' TO SELECT SETTING CAL 1,2,3,4 (HIGH/LOW/GO WILL ROTATE SETTINGS) PUT calibration weight ON (200LBS.) AND WAIT FOR STABLE LIGHT PRESS "REZERO" "CAL END" IS DISPLAYED PUT SWITCH BACK

weight scales

Acculab

ACCULAB S50 AND S100

Calibration Procedure for ACCULAB S50 AND S100 Balance

PRESS MODE TO "KG" MODE PRESS AND HOLD "CAL" UNTIL CAL. WT. MODE APPEARS PRESS "TARE" (WHILE CAL IS STILL FLASHING) "CAL O" WILL BE DISPLAYED PAUSE 2 SECONDS PLACE CORRECT CAL. calibration weight ON SCALE "CAL F" DONE WHEN DISPLAY READS CORRECT calibration weight PRESS "TARE" TO RESET ZERO

weight scales

Acculab

Acculab VA8000

Calibration Procedure for Acculab VA8000 Balance

Hold F1 & on /off self test -c- press tare f1 cal displayed put on calibration weight 5000g turn off

weight scales

Acculab

Acculab Balance

Calibration Procedure for Acculab Balance Balance

Empty Scale Press and hold Cal/mode until calibration calibration weight appears in display Put calibration weight on scale Wait until a beep. Tthe cal calibration weight will blink and it will go back into weigh mode

weight scales

Adam Equipment

TFW 60

Calibration Procedure for TFW 60 Bench Scale

Turn the Power OffTurn the power back on , during the counting from 9 to 0 press func/c key. The display will show "CAL" for a few seconds. While it is showing "CAL" press the SMPL,PRINT and TARE keys in sequence to enter the calibration section. The display will show "unload". Remove any calibration weight from the platform. Press ZERO/ENTER key. The display will show the last calibration calibration weight used. If this is correct you cn continue by pressing the ZERO/ENTER key. If it is not correct use the arrow keys to change the calibration calibration weight value. When it is correct press the ZERO/ENTER key. If the calibration is acceptable the displaywill show "PASS" and then retur to normal. If an error message is shown try calibration again as a disturbance may have prevented a successful calibration.

weight scales

Cardinal

Cardinal 2235

Calibration Procedure for Cardinal 2235 Counting Scale

Remove right side plate& throw switch Press Scale Select for scale being calibrated Change Calib)? to 1 Press count button till Load___ Load Scale Enter calibration calibration weight Press Count Unload Scale Press Count Till Done

weight scales

Denver

Denver 3000D

Calibration Procedure for Denver 3000D Balance

hold tare till cal put on 500 or 1000 or 2000 or 3000 in gram calibration weights cal x cal done

weight scales

Detecto

Detecto 2225

Calibration Procedure for Detecto 2225 Counting Scale

Turn Scale Off Remove right cover plate Remove blue Jumper in the corner of the scale attached to the display board (don't confuse with jumper on main board) Power on scale 7670? press1 then count cap= (whatever) choose remote or local now if remote just press count now. If local press scale select then count Press count button till Load___ Load Scale Enter calibration weight Press Count Unload Scale Press Count Till Done

weight scales

Digi

Digi DI-20 & 28

Calibration Procedure for Digi DI-20 & 28 calibration weight Scale

For the 20 the Jumper JP-2 must be enabled on the main board Press And hold the rezero button and press the up button to the right of the tare button then tare twice It then displays the scale increments change with arrow keys or accept and push rezero Enter the capacity of the scale push rezero Enter the calibration weight you will use to cal push rezero cal 0 appears push rezero with nothing on scale cal sp appears add calibration weight and press rezero then press tare to get out

weight scales

Digi

Digi DC-120,130 ,190

Calibration Procedure for Digi DC-120,130 ,190 Counting Scale

select scale to calibrate using" scale" button Press and hold the rezero key and press 8715 let go of rezero empty scale press rezero put calibration weight on scale enter amount of cal wt press "*" press mode twice do this per scale attached

weight scales

Digi

Digi DC-150

Calibration Procedure for Digi DC-150 Counting Scale

Press mode repeatedly until maintenance. Press and hold rezero then press 8715. release rezero press tare (right arrow key) till span adjust press enter the left of the display is the scale number to change this press + to the right of that is the scale capacity. To the riqht of that is the calibration calibration weight. The calibration calibration weight has to be entered with the same amount of decimals as the scale capacity. To change either number use the tare to highiite then enter new ualue. Once all values here are ok press enter. Empty scale push rezero add calibration weight push enter * press mode till operation press enter

weight scales

Digi

Diqi DC-60

Calibration Procedure for Diqi DC-60 Counting Scale

Remove label casette In the back of opening there is a 3 position switch. Move this switch into the middle position. Select scale to calibrate using scale button. Turn key to program Press and hold rezero while pressing 8715 release rezero. Empty scale push rezero add calibration weight enter calibration weight push enter turn key back to operate move switch back to original position

weight scales

Doran

Doran 8000

Calibration Procedure for Doran 8000 Bench Scale

Doran 8000 Remove back cover plate Throw red switch inside Displays AUGAI Press "conv" Display ENT 0 Put calibration weight on scale zero=up tare = down to correct calibration weight Press "conv" displays ent sp Take calibration weight off flip switch back

weight scales

Doran

Doran 8000XL

Calibration Procedure for Doran 8000XL Bench Scale

Doran 8000 xl press cal sw if cap ok press units cal 0 press zero put calibration weight on scale press units press print hold print to go lower or capicity to go higher till cal calibration weight is matched press zero done

Doran

4200

Calibration Procedure for 4200 Bench Scale

remove small plate in back of scale. thow red switch press gross net c000.00 press zero press units add calibration weight use arrows to correct units to save gross net

weight scales

DORAN 7000

DORAN

Calibration Procedure for DORAN Bench Scale

OPEN UP SQUARE IN BACK (RECTANGULAR) "FLIP RED SWITCH" "CO" PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT AMOUNT ON PRESS "REZERO" (COUNTS DOWN) C=calibration weight VALUE PUT THAT calibration weight ON PRESS "REZERO" (COUNTS DOWN) DONE FLIP SWITCH BACK

weight scales

Fairbanks

Fairbanks HR-2300-2

Calibration Procedure for Fairbanks HR-2300-2 calibration weight Scale

Fairbanks HR-2300-2 Press Program press ^ till cal code is 113 then enter wait for scale I forgot from there

weight scales

Fairbanks

Fairbanks UPS

Calibration Procedure for Fairbanks UPS calibration weight Scale

Fairbanks ups scale remove plate on left side press button press zero press units place displayed calibration weight on scale press cal button

weight scales

Fairbanks

Eclipse PPI-HR5000-X

Calibration Procedure for Eclipse PPI-HR5000-X Bench Scale

remove back cover press black button on board twice to get to CC CAL Hit units when reads 0 add calibration weight. Hit units again. Adjust to correct calibration weight using zero/AT sw on side of unit then hit units again

weight scales

FIling

Filing DR-2100A

Calibration Procedure for Filing DR-2100A Fork Truck Scale

Filing DR2100A 9090 ENTER DEL TOGGLES Y N MENU ENTER FOLLOW INST IF CAL SW OFF OPEN BOX AND THROW SWITCH

weight scales

Filing

Filing DR-5000

Calibration Procedure for Filing DR-5000 Fork Truck Scale

Filing DR-5000 The F1 & F2 buttons are shut off points for filling applications Cal Procedure To get into calibration mode press "9","9","Re-Zero" This calls the true zero. Go to the pot labeled R-40 and get as close as possibleto zero Push zero button Apply load Go to R-41 and dial to correct calibration weight Remove calibration weight Press "Re-zero" Re adjust if needed Once that is accomplished Call up true zero again. Go to R-40 and turn to negative 1/2 the scale capacity (-2500 to -2600lb) then rezero with button re apply calibration weight if incorrect, turn fine span pot 1/4 clockwise for up or 1/4 counter clock wise for down Do not watch the indcator because this will screw you up. Re zero the scale and re apply the calibration weight. repeat the procedure until it is correct.

weight scales

Flexweigh

DWM IV

Calibration Procedure for DWM IV Indicator

Throw SW1 ->FSE F1 Push 5 FSE F6 Press Print F6.0 0 Print Print Print f6.3d Tareprint F6.40 print Escale print capacity shows, press enter add load after countdown add load push print enter calibration weight push enter

weight scales

GSC Scale

GSC Scale

Calibration Procedure for GSC Scale Indicator

Turn On, press ¨Stanby¨, after press¨ZERO¨hold and press ¨STANBY¨Display will ready ¨SYS¨. Press ¨ZERO¨, display will ready 00000, the las digit will be blinked, Press ¨P¨ to increase to 6, press ¨NET/GROSS¨ to move to the left, press ¨P¨to increase to 5, press ¨NET/GROSS¨ to move to the left, press ¨P¨ to increase to 4, the display will read 00456, Press ¨ZERO¨, the display will show F1, press ¨SET¨,the display will show ¨F0¨, press ¨SET¨, the display will show CAL, press ¨ZERO¨, remove load from the balance, press ¨NET¨, Press ¨P¨ to include the value of weigths that will be use for calibration. Put Load on the scale and press¨ZERO¨, wait just a moment. DONE.

weight scales

GSE

GSE-350

Calibration Procedure for GSE-350 calibration weight Scale

GSE 350 Indicator Press zero and select simultaneously press zero,print,units and tare one after the other scale should say first zero. Press tare with nothing on the scale use th print and units keys to enter the load calibration weight press tare Add calibration weight press tare press tare till out

weight scales

GSE

GSE 450 460

Calibration Procedure for GSE 450 460 calibration weight Scale

GSE 450 & 460 Indicator 1) Press [Zero]+[Select] Simultaneously 2) Press [Zero] [PRINT] [UNITS] [ENTER] In that order 3) Remove any extraneous Load 4) Press[Select] until "New Zero" is dislayed 5) Press[Enter] 6) place cal calibration weight on scale 7) Key in the calibration weight value 8) Press [Enter] "Cal OK" 9) Press [Enter] 3 times

weight scales

GSE

GSE 455 465 550

Calibration Procedure for GSE 455 465 550 calibration weight Scale

GSE 455-465-550-574-650-660 1) From the calibration weight screen keyy in 100 [Select] DIsplays Key in Code: The access code must be keyed in within 5 seconds 2) Press 54321 [ID] [Enter] 3) Press Enter 4) Apply cal calibration weight 5) Key in calibration weight and press enter 6) Cal OK Press [Enter] 7) Enter=Save press [Enter] 8) Enter=Exit [Enter]

weight scales

GSE

GSE 625 622

Calibration Procedure for GSE 625 622 calibration weight Scale

GSE 625 or 622 Counting Scale Use the scale select key to choose the scale you wish to calibrate Press the [GSE][MODE] to put the instrument into offset display mode press the [GSE][ZERO]. The scale is now in the auto cal mode Before going any further enter the program jumper bypass code [23640][GSE][.] The lower display will prompt you through the set up of Auto Cal. First of all remove all calibration weight and press [Zero]. The scale will average the zero while it counts to 10 in the lower display. Counter will reseet when motion is sensed When tho count reaches 10 place the calibration weight on scale,enter that amount and press[ENTER] When the procedure is complete press[MODE] to return to offset displayed If you want to cal another scale use scale select and start all over or press [ZERO] to re-enter the Weigh Mode

weight scales

Masstron

Masstron M -1500

Calibration Procedure for Masstron M -1500 calibration weight Scale

Masstron M1500 Open Indicator and there is a fine span pot labeled R-17 Little black pot in middle of board Medium span SW3 M5000 Open indicator and main group of boards is on a hinge. undo screw in back so boards swing out. Then throw SW!-1 on the back board. then on keypad type "access" "0" "5"then press enter this is the zero cal. astrange number is there press enter till done. For span press "access" "0" "4" then enter. place calibration weights on scale. tell the indicator what it is then press enter. To save changes press access then enter twice then wait two seconds before throwing SW1-1 back.

weight scales

NCI

NCI Counting Scale

Calibration Procedure for NCI Counting Scale Counting Scale

NCI Counting scale Go to scale to calibrate Take off cover remove small cover plate throw sw2 on s2 to closed put on 1 calibration weight enter value and press acc+ Do 2nd value the same way then throw back switch

weight scales

Ohaus

Ohaus DS-10

Calibration Procedure for Ohaus DS-10 Balance

Ohaus DS-10 110 LB Capacity Hold button for 15-20 sec 0 Cal in window place 110 LB on scale it will say 50 KG Remove calibration weight

weight scales

Pennsylvania

Pennsylvania 2720

Calibration Procedure for Pennsylvania 2720 Counting Scale

PENNSYLVANIA 2720 20 LB CAPACITY SCALE: PRESS "2" PRESS "ENTER" ADD .5 THEN PUT ON 10 LBS. PRESS "ENTER" ADD 1 THEN PUT ON 20 LBS. PRESS "ENTER" ............GOES BLANK THEN TAKE calibration weight OFF PRESS (KEY IN) "20" ALL ZEROES INCLUDED PRESS "ENTER" PRESS ".005" PRESS "ENTER" (*SAME BELOW FOR SCALE 2) PRESS "2.20462" PRESS "ENTER" TURN KEY BACK SCALE 2 100 LB. CAPACITY: PRESS "2" PRESS "ENTER" ADD .5 PUT ON 50 LBS THEN "ENTER" ADD 1 PUT ON 100LBS. THEN "ENTER" .......... TAKE calibration weight OFF PRESS "100.00" THEN "ENTER" GO BACK TO *

weight scales

Pennsylvania

Pennsylvania 420-10

Calibration Procedure for Pennsylvania 420-10 Counting Scale

PENNSYLVANIA 420-10 (10/200 CAP) SAMPLE=SMALL PLATFORM BULK=LARGE PLATFORM TURN KEY TO PROGRAM KEY IN SENSITIVITY OF LOAD CELL "2" THEN "ENTER" ADD .5 (1/2 CAP. OR KEY IN OTHER calibration weight) PUT calibration weight ON PRESS "ENTER" ADD 1 (FULL CAPACITY OR ENTER IN AMOUNT) PUT calibration weight ON PRESS "ENTER" .......WILL STROBE AND THEN BLANK OUT TAKE calibration weight OFF .......WILL STROBE AGAIN AND THEN BLANK OUT (POT SCREW TO 1.5 MV OR LIFT UP CORNOR OF PLATFORM) KEY IN CAPACITY OF LOAD CELL INCLUDING ALL ZEROS (2OO.OO) PRESS "ENTER" BLANK OUT .05 RESOLUTION KEY IN PRESS "ENTER" ONLY FOR SAMPLE SCALE: PRESS "2.20462" PRESS "ENTER"

weight scales

Pennsylvania

Pennsylvania 5400

Calibration Procedure for Pennsylvania 5400 Counting Scale

Pennsylvania 5400 Counting 1)Turn Key in Back of scale 2) Enter scale number to be calibrated 3) Press enter till cal 18 STOP 4) Wait about 10 sec 5) Press Enter 6) Press enter again 7) Empty scale and press enter 8) Put 1/2 scale capacity on scale and press enter OR put different calibration weight on scale and enter value then press enter 9) Put Full scale capacity calibration weight on scale or enter value and press enter 10) Press eter till STOP 11) Turn Key Bac k

weight scales

Pennsylvania

Pennsylvania 5600

Calibration Procedure for Pennsylvania 5600 calibration weight Scale

Penn 5600 hold down 4 key full capicity press enter till cal type use units to change to load cell press enter follow instructions or enter different calibration weight values followed by enter

weight scales

Pennsylvania

Pennsylvania 7500

Calibration Procedure for Pennsylvania 7500 Bench Scale

Penn 7500 push button under plate under platter. The plate is held in place by a screw on the bottom of the scale. Push the button under the cover. Press print till cal 42 press print to cal zero. Put on half capacity and press print. Cal 44 put on full capacity and press print. Cal zero print . then print to cal 99 and the scale will count down and all done.

weight scales

Rice Lake

Rice Lake CW-40

Calibration Procedure for Rice Lake CW-40 Bench Scale

Rice Lake CW-40 Flip Red Switch behind access plate Press under button Display C press zero for zero cal scale says ent zero add calibration weight and adjust with up and down arrows. Press Units Flip switch back

weight scales

Rice Lake

Rice Lake CW-80

Calibration procedure for Rice Lake CW-80 Bench Scale

Press config Switch

Right over to calibration

S croll down to zero

C lear the platter and press the zero

witll display zcal for 10 sec

when done calsp

display shifts to full scale cap weight

adjust weight and press zero

scal is displayed while working

CalIB means it is done

remove weights

Press config switch

you are done

Rice Lake

Rice Lake IQ-600HB

Calibration Procedure for Rice Lake IQ-600HB Weigh Scale

Rice Lake IQ 600 HB TAKE DOWN METAL BAR IN FRONT 4 SWITCHES FLIP SWITCH 3 1234 (GET READING AS CLOSE TO ZERO W/PENTIAMETER) FLIP SWITCH BACK PRESS "ZERO" CLOSE SWITCH #1 (MAY NOT BE ZERO THAT`S OK) PRESS "ZERO" WITH NO calibration weight ON "CAL O" "C" PLACE 10% ON SCALE PRESS (KEY IN) 500 LBS. (NET/GROSS =UP TARE=DOWN) PRESS "CONV" FLIP SWITCH BACK

weight scales

Rice Lake

IQ 6500 6200 9500

Calibration Procedure for IQ 6500 6200 9500 Counting Scale

Rice ake IQ 6200, 6500, 9500 Press and hold the rezero key and press 8715 let go of rezero press rezero put calibration weight on scale enter amount of cal wt press "*" press mode twice do this per scale attached

Rice Lake

Rice Lake RS-130

Calibration Procedure for Rice Lake RS-130 click here

Sartorius

CP3000

Calibration Procedure for CP3000 Balance

Hold Down Tare Apply calibration weight Displayed Follow Instructions

weight scales

Scientech

Scientech

Calibration Procedure for Scientech Bench Scale

Scientech Balance 1) Press the FUNCTION switch (or MDE and RANGE switches simultaneously for dual range models) 2) Press the MODE switch when CAL1 appears. 3) Wait 10 seconds. Then press the ZERO switch 4) Place one of the flashing calibration calibration weights on the pan 5) ait 10 seconds. Then press the zero switch 6) Remove calibration weight

weight scales

Setra

Setra

Calibration Procedure for Setra Counting Scale

Setra counting scales Press "-" press"clear" press "zero" put calibration weight on scale enter calibration weight value press "set up/cal"

weight scales

Sterling

Sterling

Calibration Procedure for Sterling Counting Scale

Sterling 825 Counting Scale 1)Press Scale Select to go to the scale you are calibrating 2) Press and hold Gross/Net Key until Params 3) Now the NTEP parameter along with its value is displayed 4)Press the print key or zero key till FCAL 5)Select FCALT with tare key and press Print 6) Shows current number f scales to change tis pram press tare or press print to move on 7) This is capacity change it or press print to move on 8)Graduations press print or change 9)Shows Cal Units print to move on 10) CAL F Change this to CAL T using Tare button press print to begin cal a)Clear scale and press print b)Step A8 press print c)Place calibration weight on scale, enter calibration weight value and press print 11)SPN 0.0 this is the span error press print 12)Shows caacity of next scale. Repeat steps 7-11 13)Gross/Net or reset key will get you out at the end

weight scales

Toledo

Toledo 8571

Calibration Procedure for Toledo 8571 Counting Scale

TOLEDO 8571 PRESS "ENTER" UNTIL E SCALE (F 3.2 F 3.3) AFTER E SCALE PRESS "*" (PRINT) ADD LD (TARE=CHANGE #'S CLEAR=NEXT # ZERO=BACKS UP) "OOO" ADD calibration weight PRESS "*" (PRINT) PRESS "CLEAR" PRESS "*PRINT" UNTIL F 4.5 THEN "*" (PRINT) SET OFF

weight scales

Toledo

Toledo 8581

Calibration Procedure for Toledo 8581 calibration weight Scale

Toledo 8581 Counting Scale Throw Switch under platter Press 0 to F3 Press print to F3.2 Press1 Scale Cap Displayed Press Print Increment Size Press Print E Scl (Empty Scale) Press Print Scale Counts Down Add calibration weight Press Print Enter calibration weight in whole numbers Press Print Counts Down Cal d (Done) Press print till set up off Throw switch back Turn scale off & on

weight scales

Toledo

Toledo SC-05

Calibration Procedure for Toledo SC-05 calibration weight Scale

Toledo SC-05 lift platter. Press cal switch in hole to the right of platter. Puts yuu into service. Use rezero enter and tare to get into cal mode

weight scales

Transcell

TI-500E

Calibration Procedure for TI-500E Scale Indicator

Turn Off Scale. Throw Switch on back of scale. Turn On Scale. F1 press right arrow till F16. Press Down Arrow. C0 Press Zero Key. Press Net/Gross. Says Set. F17 with right arrows. Arrow down. C1 then flash 0. Add calibration weight. Change weigh using arrow buttons. Press Gross/Net end C1 Turn off and throw switch back

weight scales

Vibra Screw

Filling System

Calibration Procedure for Filling System Filling System

Throw cal switch on dip switches located at the bottom of aluminum box. Press mode to get to gross mode press the 8 button (also says cal). Displays SPCd press enter enter 5321 then enter setup press enter till cal press enter Hd Cal Press enter "Zero" displays press enter "-" displays press test clr 0.0 displays press enter span displays place calibration weight on scale press enter "-" displays press test/clr enter calibration weight press enter lincor press enter 0.0 displays press enter return displays press enter end cal displays press enter switch back

weight scales